6 data formats, Character is, A, k, p, r, i, n, t – Rice Lake 720i Programmable Indicator/Controller - Installation Manual User Manual

Page 112

Advertising
background image

106

720i Installation Manual

11.6

Data Formats

Continuous Output Serial Data Format

If continuous transmission is configured for a serial
port (STREAM parameter set to LFT or INDUST on
the SERIAL menu), the

720i

sends data using the

Consolidated Controls serial data format shown in
Figure 11-1.

<STX> <POL> <wwwwwww> <UNIT> <G/N> <S> <TERM>

ASCII 02

decimal

Polarity:
<space> = Positive

<–> = Negative

Weight: 7 digits, right-justified, dummy

zeroes, decimal point with no leading
zeroes except for leading zero immediately
preceding the decimal point. Leading

zeroes transmitted as spaces.

L = pounds
K = kilograms
T = tons
G = grains
<space> = grams

O = ounces

G = Gross

N = Net

Status:
<space> = valid

I = Invalid
M = Motion
O = Over/under range
Z = COZ

<CR> <LF>

or <CR>

Figure 11-1. Continuous Output Serial Data Format

Demand Output Serial Data Format

When demand mode is configured for the serial port
(STREAM parameter set to OFF), the

720i

uses a data

string formatted for a basic ticket printout. The
particular ticket format printed depends on the
indicator configuration.

You can customize the ticket to work with a wide
variety of printers, scoreboard displays, and other
remote equipment. See Section 6.0 on page 52 for
more information on custom print formats.

RS-485 Data Formats

Four-wire RS-485 communications is supported on the
“A” ports of any installed serial expansion card.

The RS-485 software protocol is enabled when you
assign a non-zero address to the indicator. Valid
RS-485 addresses must be in the range 1–255; the
address is specified on the ADDRESS parameter on the
SERIAL menu.

All remote commands are initiated using the data
format shown in Figure 11-2.

<STX> <ADDRESS> <COMMAND> <CR>

ASCII 02

decimal

Address of the

r eceiving indicator

EDP serial command

NOTE: Host must send <CR>, not

<CR><LF>. Failure to use <CR>

renders all indicators unable to
respond to serial commands

ASCII

13 decimal

Figure 11-2. RS-485 Send Data Format

If the initiating device address matches the port address
of an

720i

on the RS-485 network, that indicator

responds. For example, with demand outputs, or in
response to a KPRINT command, the responding
indicator uses the format shown in Figure 11-3.

<STX> <ADDRESS> <DEMAND DATA> <ETX> <CR>

ASCII 02

decimal

Addr ess of the

transmitting indicator

Response commands from
indicator:
< first line > <EOL>
< additional lines > <EOL>
< last line > <EOL>
where EOL = <CR> or <CR> <LF>

(set by TERMIN parameter on the
SERIAL menu)

ASCII 13

decimal

ASCII 03

decimal

Figure 11-3. RS-485 Respond Data Format

Example:

To send the KPRINT command from an

ASCII terminal to an indicator at address 65 (decimal)
on the RS-485 network, use the format shown in
Figure 11-2.

•

The keyboard equivalent for the start-of-text
(

STX

) character is

CONTROL-B

.

•

The indicator address (65) is represented by an
upper case “A”.

•

The carriage return (CR) character is generated
by pressing the

Enter

key.

Therefore, to send the KPRINT command to the
indicator at address 65, enter the following at the
terminal:

CONTROL-B

, A, K, P, R, I, N, T,

ENTER

.

The indicator responds with the format shown in
Figure 11-3:

<STX> A

SCALE #1

<EOL>

GROSS 1699 LB<EOL>
01/14/2008 10:05 AM<EOL>

<ETX> <CR>