Serial Commands
87
10.1.2
Reporting Commands
Reporting commands send specific information to the
serial port. The commands listed in Table 10-2 can be
used in both setup mode and normal mode.
Table 10-2. Reporting Commands
Command
Function
DUMPALL
List all parameter values
DUMPAUDIT
List audit trail information
SPDUMP
PCEE
configuration
VERSION
Write
720i
software version
HARDWARE
Lists installed option cards. See
Section 11.1.2 on page 99 for more
information about using the HARDWARE
command.
XE
Returns a 10-digit code representing any
error conditions currently shown on the
front panel. See Section 11.1.5 on
page 100 for more information.
10.1.3
Clear and Reset Commands
The following commands can be used to clear and reset
the
720i
:
RS
: Reset system. Resets the indicator without
resetting the configuration.
RESETCONFIGURATION
: Restores all configuration
parameters to their default values (setup mode
only). The
RESETCONFIGURATION
function can
also be initiated by pressing the
Reset Config
softkey under the VERSION menu. Use the
up
key
to select
Clear entire indicator configuration
, then press
Enter
to reset the indicator.
Note
All load cell calibration settings are lost when
the RESETCONFIGURATION command is run.
10.1.4
Parameter Setting Commands
Parameter setting commands allow you to display or
change the current value for a particular configuration
parameter (Tables 10-3 through 10-10).
Current configuration parameter settings can be
displayed in either setup mode or normal mode using
the following syntax:
command<ENTER>
Most parameter values can be changed in setup mode
only.
Use the following command syntax when changing
parameter values:
command=value<ENTER>
, where
value
is either a number or a parameter value. Use no spaces
before or after the equal (=) sign. If you type an
incorrect command, the display reads
??
.
For example, to set the motion band parameter on Scale
#1 to 5 divisions, type the following:
SC.MOTBAND#1=5<ENTER>
For parameters with selectable values, enter the
command and equal sign followed by a question mark:
command=?<ENTER>
to see a list of those values. The
indicator must be in setup mode to use this function.
Note
Some parameters are valid only if other
parameters or parameter values are specified.
See the configuration menus in Section 3.2 on
page 20 for information about parameter
dependencies. Restrictions for front-panel
configuration also apply to serial command
configuration.
Command
Description
Values
SC.GRADS#1
Graduations
1–9999999
SC.SPLIT#1
Multi-range or multi-interval scale type
OFF, 2RNG, 3RNG, 2INTVL, 3INTVL
SC.ZTRKBND#1
Zero track band
0, 0–100
SC.ZRANGE#1
Zero range
1.900000, 0–100
SC.MOTBAND#1
Motion band
1, 0–100
SC.SSTIME#1
Standstill time
1–65535
SC.OVRLOAD#1
Overload
FS+2%, FS+1D, FS+9D, FS
SC.WMTTHRH#1
Weighment threshold
grads
SC.NUMWEIGH#1
Number of weighments
—
SC.MAX_WEIGHT#1
Maximum weight
—
SC.DIGFLTR1#1
SC.DIGFLTR2#1
SC.DIGFLTR3#1
Digital filtering
1, 2, 4, 8, 16, 32, 64, 128, 256
SC.DFSENS#1
Digital filter cutout sensitivity
2OUT, 4OUT, 8OUT, 16OUT, 32OUT, 64OUT, 128OUT
Table 10-3. SCALES Serial Commands