Rice Lake 720i Programmable Indicator/Controller - Installation Manual User Manual
Page 41

Configuration
35
SERIAL Menu
Parameter
Choices
Description
Level 2 submenus
PORT 2
PORT 4
CMD
SCALE
IND SC
Specifies the type of data received by the port:
SCALE:
Legal-for-trade serial scale input
IND SC:
Industrial (non-legal-for-trade) scale input
KEYBD is available only on Port 1; SCALE, IND SC are available only on Port 4 and higher
(expansion ports).
Note
The keyboard interface is not hot-pluggable. Power-off the
720i
before attaching keyboard cable to the Port 1 connector.
DO NOT EDIT PORT 3 PARAMETERS. Port 3 provides communications
with the detached display module (DDM); changing these parameters will disable the
display.
Print format outputs default to Port 4; if using Port 4 for other scale connections,
reassign needed print formats to another port before configuring.
Level 3 Submenus
Ports 2 and 4
BAUD
9600
19200
28800
38400
57600
115200
300
600
1200
2400
4800
Baud rate. Selects the transmission speed for the port.
Note
The maximum baud rate for ports on serial expansion cards
(port number greater than 4) is 19200.
BITS
8NONE
7EVEN
7ODD
8ODD
8EVEN
Selects number of data bits and parity of data transmitted or received by the port.
STOP BITS
2
1
Selects number of stop bits transmitted or received by the port.
ECHO
ON
OFF
Specifies whether characters received by the port are echoed back to the sending unit.
RESPONSE
ON
OFF
Specifies whether the port transmits replies to serial commands.
TERMIN
CR/LF
CR
Termination character. Selects termination character for data sent from the port.
EOLDLY
0
0–255
End-of-line delay. Sets the delay period, in 0.1-second intervals, from when a formatted line is
terminated to the beginning of the next formatted serial output. Value specified must be in the
range 0-255, in tenths of a second (10 = .1 second).
HANDSHK
OFF
XONOFF
HRDWAR
Specifies whether XON/XOFF flow control characters or hardware handshaking is used.
Hardware handshaking is available only on Port 2.
PORTTYPE
232
485
Specifies whether the port is used for RS-232 or RS-485 communications. If 485 is selected
(Port 7 only), additional prompts are shown to specify half- or full-duplex operation and
RS-485 address.
Note
RS-485 communications can be specified for Port 7 (serial expansion card)
only.
Table 3-6. Serial Menu Parameters