Appendix
109
11.9
PS/2 Keyboard Interface
Serial port 1 (connector J3) on the
720i
CPU board
provides a PS/2-type keyboard interface (Scan Code
S e t 1 o n l y ) f o r u s e w i t h a r e m o t e k e y b o a r d .
Table 11-10 shows the pin assignments for a PS/2
keyboard cable.
Table 11-10. PS/2 Keyboard Connector Pin Assignments
Pin
Signal
Male Connector
1
DATA
1
2
3 4
5
6
2
(reserved)
3
GND
4
+5V
5
CLOCK
6
(reserved)
720i
-specific functions
provided by the keyboard interface; most other
alphanumeric and navigational keys provide functions
equivalent to those typical for PC operation. Menu
parameters and serial commands that affect indicator
k e y p a d o p e r a t i o n ( i n c l u d i n g t h e K B D L C K ,
ZERONLY, and KLOCK serial commands) also affect
the remote keyboard.
Note
The keyboard interface is not hot-pluggable.
Disconnect power to the
720i
before
connecting the keyboard cable to the Port 1
connector.
Table 11-11. PS/2 Keyboard Functions
Key
Function
F1
Softkey 1
F2
Softkey 2
F3
Softkey 3
F4
Softkey 4
F5
Softkey 5
F6 (Alt+Z)
ZERO key
F7 (Alt+G)
GROSS/NET key
F8 (Alt+T)
TARE key
F9 (Alt+U)
UNITS key
F10 (Alt+P)
PRINT key
F11
Not used
F12
Print Screen
Same as PRINT key, in both
normal and setup modes
11.10 Audit Trail Support
Audit trail support provides tracking information for
configuration and calibration events. To prevent
potential misuse, all configuration and calibration
changes are counted as change events.
Use
Revolution III
to display audit trail information.
The audit trail display includes the legally relevant
(LR) version number (software version for the code
that provides audit trail information), a calibration
count and, if REGULAT=NTEP, a configuration count.
The exact format of the information shown depends on
the regulatory agency specified for the REGULAT
(FEATURE menu) parameter.
Audit trail information can be printed from
Revolution
III
or by sending the DUMPAUDIT serial command.
By pressing the
MENU
key , the Audit Trail menu is
bold. Pressing the
Enter
key will display the LR
software number and the current audit trail counter.
When done, press
Enter
to return to weighing mode.
For NTEP and Measure Canada modes
, there are two
numbers displayed; Configs and Calibs. The Configs
n um be r t r a ck s c ha ng es t o the legally relevant
configuration parameter changes. The Calibs number
tracks changes to the scale calibration.
In OIML mode
, there is only one number displayed;
Audit Trail. The Audit trail number is incremented for
any legally relevant configuration parameter changes
and all calibration changes. The last date the scale was
calibrated is also displayed.
long tons
2240.00
pounds
1016.05
kilograms
1.12000
short tons
1.01605
metric tons
troy ounces
480
grains
31.10348
grams
0.031103
kilograms
1.09714
ounces
0.068571
pounds
0.083333
troy pounds
troy pounds
5760
grains
373.2417
grams
0.373242
kilograms
13.16571
ounces
0.822857
pounds
12
troy ounces
Primary Unit
x Multiplier
Secondary/
Tertiary Unit
Table 11-9. Conversion Factors (Continued)