Key-press route to acuity menu, Press net off to disconnect from acuity, Button as described below. the – Welch Allyn 20хEL Propaq Encore Vital Signs Monitor - Reference Guide User Manual

Page 117

Advertising
background image

Press NET OFF to Disconnect from Acuity

Propaq Encore Reference Guide

117

Welch Allyn

Key-press Route to Acuity Menu

Press NET OFF to Disconnect from Acuity

When you want to disconnect the monitor from Acuity, be sure to use the

NET OFF

button as

described below. The

NET OFF

button is a safety feature to help make sure the patient is not

disconnected accidentally. If you simply disconnect the Acuity network cable from the
monitor without using the

NET OFF

button, the monitor and Acuity both generate equipment

alert messages that must be acknowledged by an operator.

1.

To disconnect the Propaq Encore from the Acuity network, press the front panel

NET

OFF

button.

Warning If you don’t set alarm limits, Acuity uses preset settings (for
arrhythmia limits), and the powerup default settings for the Propaq Encore.

When a Propaq Encore in Adult or Pediatric Mode is connected to an Acuity
System, the audio alarms at the bedside Propaq can be delayed up to 4
minutes and 15 seconds. The delay time is selected in Acuity software at the
time of Acuity installation. Visual alarm indications are not delayed.

FRZ PRNT

SETUP

NIBP ECG/RESP

INV

PRS

SpO2/CO2

SETUP MENU 1

MORE

ALARMS

TRENDS

WAVE SEL

STATSCALE

ACUITY

ACUITY MENU

PREV MENU

SNAPSHOT

NET OFF

MAIN MENU

FREEZE/
UNFREEZE

NET OFF

ART

( 85)

122

58

HR

80

NET OFF

When the monitor is connected to Acuity, pressing

this button disconnects the monitor from the

Acuity network.

FREEZE/UNFREEZE

When the monitor is NOT connected to Acuity,

pressing this button freezes or unfreezes the

displayed waveforms.