Connect to a Non-Intubated Patient

1.

Position the cannula on the patient according to the manufacturer’s instructions.

2.

Connect the sample line to the cannula and the
watertrap. Make sure that the sample line is
firmly connected.

Warning The cannula is disposable and should only be used for a single
patient. Do not reuse the cannula for another patient.

If oxygen is being delivered while using Sidestream

CO

2

, be sure to use a

CO2

Sampling and O2 Delivery Cannula. Using a different type of cannula could
obstruct oxygen delivery.

The exhaust port for Sidestream

CO

2

is an output for the expired gases from

the patient and any connected breathing apparatus. The exhaust port is
intended only for connection to gas collection equipment such as gas
scavenger devices. Do not allow any other connection to the exhaust port.

If you use a gas scavenging system with Sidestream

CO

2

, be sure to install it

according to the manufacturer’s instructions. The scavenging system should
comply with ISO 8835-3:1997 (E).

Sidestream

CO

2

accuracy decreases if

additional tubing is connected to the
sample line. Avoid connecting additional
tubing to the standard sample line.

Sample line

Watertrap

WARNING - Do not connect sample

line or patient input to exhaust port.

Exhaust Port