Capnography (CO2)
Propaq Encore Reference Guide
89
Welch Allyn
OCCLUSION - CHECK WATERTRAP/TUBING
. Blockage has been detected on the
Sidestream CO
2
input. Check the watertrap, sample line, and any connected tubing for
occlusion.
PUMP FAILURE,
SERVICE REQUIRED. The pump is not able to maintain the target
flow rate. Send the Propaq to a Biomedical Engineer for service.
SIDESTREAM STICK EEPROM FAILURE
. Send the Propaq to a Biomedical Engineer
for service.
SSP BOARD EEPROM FAILURE
. Send the Propaq to a Biomedical Engineer for service.