Pulse Oximetry (SpO2)

Welch Allyn

72

Propaq Encore Reference Guide

9.

If the C-LOCK function is desired, press

C-LOCK

to set it to ON.

10.

Set alarm limits according to your hospital’s standards.

11.

If patient movement interferes with measurements, consider the following possible
solutions:

be sure the sensor is secure and properly applied

use a new sensor with fresh adhesive backing

select a different type of sensor

move the sensor to a less active site

Response

Indications for Use

NORMAL: 5-7 seconds

Use for relatively stable patients.

FAST: 2-3 seconds

Use when patient movement or other artifact is not present.

SLOW: 10-15 seconds

a

a.SLOW setting is not applicable to the Nellcor

SpO

2 option with motion tolerance.

Use when patients exhibiting movement are preventing
accurate measurement at NORMAL setting.

C-LOCK synchronizes the pulse oximeter's systole determination to the R-
wave to reduce the effects artifact may have on SpO

2

measurements. Under

some conditions you may find more stable SpO

2

readings with C-LOCK set to

ON. SYNC appears next to the waveform when synchronization to the ECG
has been obtained. Synchronization takes a few seconds to establish the first
time. If C-LOCK is on and the HR source is SpO

2

, the heart rate source is

automatically changed to ECG. An ECG signal must be present or C-LOCK
does not activate.

If you get false SpO

2

alarms with patients with low perfusion states or multiple

arrhythmias, try turning off C-LOCK.

To help minimize false alarms, the Propaq monitor briefly delays or "holds off"
triggering both audible and visual alarms for limit violations for SpO

2

% and

Pulse Rate for 10 seconds. After the alarm hold-off period begins, if the
monitor detects that the patient’s vital sign has returned to acceptable limits,
the monitor cancels the alarm hold-off. The next time a vital sign limit is
violated, the monitor starts a new hold-off period.