Welch Allyn 20хEL Propaq Encore Vital Signs Monitor - Reference Guide User Manual

Page 82

Advertising
background image

Capnography (CO2)

Welch Allyn

82

Propaq Encore Reference Guide

3.

Press

mm/s

to select either 3.13, 6.25 or 12.5 mm/sec. The default setting is 6.25 mm/

sec. This sweep speed also applies to RESP.

4.

Press

MORE

to view the CO

2

status window.

5.

If either O

2

or N

2

O is being administered to a patient, press

GAS COMP

to set the proper

gas compensation as listed below. If no other gas is being administered, set GAS
COMPENSATION to OFF. (OFF is the default setting.)

6.

Press

RESPONSE

to select either NORMAL, SLOW, or FAST.

The FAST setting is recommended where a sudden step change in ETCO

2

is of

concern, such as that induced by an air embolus in certain neurosurgical procedures.

A SLOW response will decrease ETCO

2

false alarms when breath morphology varies

considerably from one breath to the next.

The default setting is NORMAL.

z

7.

Set alarm limits for ETCO

2

, INCO

2

, and Breath Rate.

If an inspired value is displayed indicating patient rebreathing (non-zero
IN

CO

2

), check the patient breathing circuit for proper function. Verify the

sensor calibration against room air. If the Propaq continues to display inspired
values, return the sensor to Welch Allyn for service.

Gas Administration Level/GAS

COMP Setting

ETCO

2

or INCO

2

Value

OFF

CO

2

value = actual

CO

2

value

O

2

> 50%, No N

2

O

CO

2

value = actual CO

2

value x 1.03

N

2

O > 50%

CO

2

value = actual CO

2

value x 0.952

If the ET

CO

2

value is displayed as + + +, verify calibration against a known

reference gas. If the sensor calibration is not accurate, return it to Welch Allyn
for service.

Response Time Setting

Sampling Period

Typical Indications for Use

FAST

15 seconds

During neuroanesthesia

NORMAL

30 seconds

During routine use

SLOW

45 seconds

To decrease ET

CO

2

false alarms

Advertising