Nellcor spo, Specifications – Welch Allyn 20хEL Propaq Encore Vital Signs Monitor - Reference Guide User Manual

Page 145

Advertising
background image

Pulse Oximetry (SpO

2

)

Propaq Encore Reference Guide

145

Welch Allyn

Nellcor SpO

2

Specifications

Characteristic

Specification

a

a.When performing SpO2 monitoring, the monitor SpO2 channel may not (in rare instances) initially detect a damaged SpO2
sensor or extension cable (as described in EN865:1997, Section 51.109), but may simply display a flat line and no numerics. If
this occurs, try connecting a new SpO2 sensor (and extension cable, if needed) and restart the monitor to resume normal
monitoring.

Saturation (% SpO

2

)

Range
Resolution
Alarm Limits

b

b.Minimum difference between upper and lower alarm limits is 2%.

0% to 100%
1%
52% to 100% (upper)
50% to 98% (lower)

Probe Accuracy

c

(saturation levels

between 70% and 100%, 28° to
42°C)

Adult/Pediatric
Neonatal

c.Refer to the Welch Allyn

Products and Accessories

guide (810-0409-XX) for accuracy specifications for all Nellcor SpO

2

probes recommended for use.

Digit accuracy: ±2 counts
Digit accuracy: ±3 counts

Pulse Rate

Range

d

Alarm Limits

d.When using the pulse oximetry option with motion tolerance, a substantial and rapid (<2 seconds) drop in pulse rate may result
in erroneous pulse rate readings and loss of the audible pulse indicator.

Motion tolerant option: 25 to 249 beats per minute
Option without motion tolerance: 25 to 250 beats per minute
27 to 250 beats per minute (upper)
25 to 248 beats per minute (lower)

Pulse Rate Accuracy

No Motion
During Motion

±3 beats per minute
±5 beats per minute

Alarm Hold-Off Time Period

10 seconds; resets if the sensor reports levels within limits before 10
seconds elapses

Circuitry

Microprocessor controlled
Automatic self-test of oximeter when powered on
Automatic setting of default parameters
Automatic alarm messages

Electrosurgery interference
suppression

Yes

Sensor Compatibility

Compatible only with Nellcor sensors listed in the Welch Allyn

Products

and Accessories

booklet.

Sensor LEDs

RED Wavelength
INFRARED (IR) Wavelength

660 nm (nominal)
890 nm (nominal)

Sensor Energies (Radiant Power)

Electrical Power
Optical Power

52.5 mW max.
15 mW max.

Pulse Rate Accuracy

No Motion
During Motion

±3 beats per minute
±5 beats per minute