Time/Day Settings and Trends

When you change the hour/minute/second setting for the monitor in the Time/Day
window, the monitor deletes any patient trend data that is older than five hours for non-
NIBP trends or older than eight hours for NIBP trends according to the new clock setting.

However, if the monitor has not yet stored the full capacity of trends and you change the
hour/minute/second setting to a time that is within the stored trend period, previously
stored trends are not erased.

Changing the day, month, or year setting does not affect the stored patient trends.

Changing the Date Format, Filter, and Units

To change the date format, ECG filter, or some measurement units, first make sure you are in
the Adult patient mode. Then press

SETUP

,

MORE, MORE

,

SERVICE, YES

(to access the Service

Menu)

, MORE, MORE, SETTINGS

. The monitor displays the Settings window:

Warning Changing the hour/minute/second setting for the monitor in the
Time/Day window can cause the monitor to erase previously stored patient
trend data.

NEXT

Selects the next setting in the status window.

CHANGE

Changes the currently selected display setting.

DATE

Sets the date format: Month/Day/Year, Day.Month.Year, or Year/
Month/Day.

FILTER

Sets the ECG filter frequency. Make sure it is set to your ac mains
frequency.

TEMP F/C

Sets the temperature display units: either degrees Fahrenheit or Celsius.
If you change the units, the TEMP trends will not be cleared.