Welch Allyn 20хEL Propaq Encore Vital Signs Monitor - Reference Guide User Manual

Page 61

Advertising
background image

NIBP

Propaq Encore Reference Guide

61

Welch Allyn

NIBP Displayed in Waveform Window

AUTO/MAN

This button switches the mode between Automatic or Manual Mode. The
Manual Mode is the default unless you change it by reprogramming your
Propaq. Measurements can be taken at intervals of 1, 2, 3, 5, 10, 15, 30, and 60
minutes. Press

START

to initiate a measurement.

INTERVAL

Selects the interval at which NIBP measurements are automatically taken. The
interval you select, ranging from one minute to 60 minutes, is shown on the
display next to the word TIME.

TURBOCUF

Automatically starts NIBP measurements and continues to take as many
measurements as possible within five minutes.

SMARTCUF

Enables or disables the Smartcuf motion artifact filter. NIBP measurements can
still be taken when Smartcuf is off. Artifact may interfere with the accuracy of
NIBP measurements with Smartcuf off.

mmHg

NIBP

SETUP

INV PRS

SpO

2

/CO

2

ECG/RESP

MCO

2

BR

HR

T

1

NIBP

18:45:28

III

1

mV

cm

MON

N

I

B

P

60
30
0

102.3

°

F

mmHg TIME SINCE READING: 0 min

35

135/ 81

(103)

12

80

NIBP

C

O

2

M

(103)

135

81

SpO

2

99

mmHg

NIBP

SETUP

INV PRS

SpO

2

/CO

2

ECG/RESP

MCO

2

BR

HR

T

1

NIBP

19:02:28

III

1

mV

cm

MON

N

I

B

P

60
30
0

102.3

°

F

SYS DIA MEAN

135 81 (103)

mmHg TIME SINCE READING: 17 min

35 12

80

NIBP

18:45

C

O

2

M

135/

81

(103)

SpO

2

99

NIBP TREND

TIME

BR

SYS/DIA-MEAN

HR/PR

BPM

Br/M

HH:MM
10:35

9:46

9:31

9:20

OFF

OFF

OFF

OFF

106

88

93

91

OFF

OFF

OFF

OFF

135/ 81 103

142/ 83 110

145/ 87 112

126/ 85 91

SpO2
%

mmHg

PR

T

1

NIBP

10:45:28

N

I

B

P

102.3

°

F

mmHg TIME SINCE READING: 10 min

106

NIBP

PRINT

OXYCRG

NXT TRND

135/ 81

(103)

(103)

135

81

The numerics are shown in large

characters for 16 minutes for each

new measurement taken...

By turning on NIBP in the wave

select window, the NIBP numerics

can be displayed in a waveform

window.

... and then they change to the

smaller characters for 44 minutes.

The numerics are removed after 60

minutes.

If NIBP is the only vital sign being

monitored, the numerics are

displayed in a waveform window

above a trend window.