Invasive Pressure

Propaq Encore Reference Guide

55

Welch Allyn

ZERO REJECTED (or REJECT).

Unable to establish a zero reference value. The message

remains until the

CANCEL

button is pressed.

CANCELED.

This message appears if

CANCEL

is pressed while the channel is zeroing.

These equipment messages can appear in an equipment alert window.

TRANSDUCER NOT DETECTED.

The transducer connection is broken.

TRANSDUCER SHORT CIRCUIT.

This message appears when the Propaq Encore

senses a short in the transducer. The transducer should be replaced.

INCOMPATIBLE TRANSDUCER

. Check the compatible transducers listed in the Welch

Allyn Products and Accessories booklet to ensure you are using a compatible transducer.