Monitor Setup

Welch Allyn

34

Propaq Encore Reference Guide

From the Main Menu, press

SETUP

,

MORE, CHANGE, SETUP

. The Mode Setup window appears:

The asterisk (*) indicates which patient mode is currently selected for powerup.

Whenever you change the patient mode, the alarm limit settings automatically
change to the settings for that mode.

NEXT

Selects the next setting in the status window.

POWERUP*

Selects the highlighted patient mode (and its associated settings) as the
powerup mode. The selected powerup mode is marked by an asterisk (*).
(This does not change the current patient mode.)

USE NOW

Selects the highlighted patient mode (and its associated settings) as the
current patient mode. (This does not affect the powerup mode.)

SAVE

Use this button to reprogram the settings of Custom patient mode:

1.

Make sure the patient mode you want to reprogram (

ADULT

,

PED

or

NEO

) is

currently-used (to change patient modes, highlight the desired mode and
press

USE NOW, YES

).

2.

Exit the Mode Setup window, then use other menus and buttons to set the
monitor settings and alarm limits as desired.

3.

Re-enter the Mode Setup window, highlight the desired Custom mode, and
press

SAVE, YES

.

BATTERY: 9.3 VOLTS

mmHg

NEXT

PREV MENU

USE NOW

POWERUP*

MCO

2

BR

MODE SETUP

FACTORY:

NO ZERO

38 12

97

*ADULT

PED

NEO

CUSTOM

ADULT

PED

NEO

* = PATIENT MODE ON POWERUP

SAVE

P

2

:

Sp02

ZEROED