Factory Default Settings

Propaq Encore Reference Guide

159

Welch Allyn

Display Wave Select

All waves are on except NIBP

Trend Group

NIBP

Alarm Limits

All are ON except P2

HR Limits

Adult: 50, 120 beats per minute
Ped: 50, 150 beats per minute
Neonate: 100, 200 beats per minute

NIBP Limits - Systolic

Adult: 75, 220 mmHg
Ped: 75, 145 mmHg
Neonate: 50, 100 mmHg

NIBP Limits - Diastolic

Adult: 35, 110 mmHg
Ped: 35, 100 mmHg
Neonate: 30, 70 mmHg

NIBP Limits - Mean

Adult: 50, 120 mmHg
Ped: 50, 110 mmHg
Neonate: 35, 80 mmHg

P1, P2 Limits - Systolic

Adult: 75, 220 mmHg
Ped: 75, 145 mmHg
Neonate: 50, 100 mmHg

P1, P2 Limits - Diastolic

Adult: 35, 110 mmHg
Ped: 35, 100 mmHg
Neonate: 30, 70 mmHg

P1, P2 Limits - Mean

Adult: 50, 120 mmHg
Ped: 50, 110 mmHg
Neonate: 35, 80 mmHg

SpO

2

Limits

Adult: 85%, 100%
Ped: 85%, 100%
Neonate: 80%, 95%

RR/BR

Adult: 5, 30 BrM
Ped: 10, 45 BrM
Neonate: 10, 60 BrM

TEMP Limits

35.0°, 37.8° C

∆T Limits

0.0°, 2.8° C

ETCO

2

Limits

25, 60 mmHg (3, 8 for % and kPa)

INCO

2

Limits

N/A, 5 mmHg (0.7 for % and kPa)

Apnea Delay

Adult/Ped: 20 seconds
Neonate: 15 seconds

PRINTER SETTINGS

Printer Alarm Print

OFF

Printer Auto Print

OFF

Printer NIBP Ticket

OFF

Printer Apnea Ticket

ON

Printer Print Speed

25 mm/sec

Printer Auto Trend

OFF

Printer Trend Selections

NIBP and P1 = ON; all others = OFF

Printer OxyCRG on Alarm

OFF

a.Any time you change the Date, Filter, Temp F/C, Decimal, or CO

2

Units setting, the new setting also becomes

the powerup default setting.