Pulse Oximetry (SpO2)

Welch Allyn

74

Propaq Encore Reference Guide

STANDBY

is displayed in the numeric window when the SpO

2

sensor is disconnected

from the patient, an alarm occurs, and you press the

STANDBY

button. STANDBY is also

displayed if you first plug the SpO

2

sensor cable into the monitor connector before

attaching the SpO

2

sensor to the patient.

DEFECTIVE SpO2 SENSOR

. If you see the error message DEFECTIVE SpO2 SENSOR,

either the sensor is not compatible with the monitor or the sensor is not working
properly. Check the Welch Allyn Products and Accessories booklet (810-0409-XX) to be
sure the sensor is compatible. If compatibility is not a problem, try another sensor.