Monitor (environmental) – Welch Allyn Vital Signs Monitor, Propaq Encore - Quick Reference Guide User Manual

Page 119

Advertising
background image

Reference Guide

Specifications

113

Monitor (environmental)

Caution The monitor may not meet performance specifications if it is not used
or stored within these environmental specifications.

Characteristic

Specification

Operating Temperature

0° to 40° C

Shipping and Storage Temperature

-20° to 60° C

Operating Altitude

-2,000 to 15,000 ft (-610 to 4,572 m)

Shipping and Storage Altitude

-2,000 to 40,000 ft (-610 to 12,192 m)

Operating Relative Humidity

15% to 95%, noncondensing per MIL STD 810E, Procedure 1-natural

Shipping and Storage Relative
Humidity

15% to 95%, noncondensing per MIL STD 810E, Procedure 1-natural

Shock

102 g

Vibration, Random

Designed to meet EN ISO 9919:2005
Designed to meet EN ISO 21647:2004

0.02g

2

/Hz from 10 to 500 Hz, ramping down to 0.002g

2

/Hz at 2000 Hz.

Operating 1 hour per axis, 3 hours per test. Designed to meet RTCA DO-
160C, Category C.
Clause 21.102b) 10Hz to 100Hz: 5.0 (m/s

2

)

2

/Hz, 100Hz to 200Hz at –7 db

per octave; 200 Hz to 2000 Hz: 1.0 (m/s

2

)

2

/Hz; duration: 30 min per

perpendicular axis (3 total)

Vibration, Sinusoidal

0.10 peak to peak inches 5 to 17 Hz, sloping to 0.01 peak to peak inches at
55 Hz, then sloping to 0.0001 peak to peak inches at 2000 Hz. Operating 1
hour per axis, 3 hours per test. Designed to meet RTCA DO-160C, Category
N.

Degree of protection against ingress
for monitors without CO

2

or printer

options

IPX1 rating, drip proof per EN60529: 1991

Advertising