Power adapters – Welch Allyn Vital Signs Monitor, Propaq Encore - Quick Reference Guide User Manual

Page 123

Advertising
background image

Reference Guide

Specifications

117

Power adapters

Characteristic

Specification

Protection classifications, all adapters

a

a.

Per EN 60601-1 unless otherwise stated.

Type of Protection Against Electric
Shock

Class I, (Protectively Earthed)

Degree of Protection Against Harmful
Ingress of Water

For ordinary, indoor locations only.

Method of Disinfection

Not suitable for autoclaving

b

b.

See Chapter 8 for cleaning instructions.

Flammable Anesthetics

Not suitable for use with flammable anesthetics

Environmental specifications, all adapters

Operating Temperature

0° to 50° C

Shipping and Storage Temperature

-20° to 60° C

Operating Altitude

-2,000 to 15,000 feet (-610 to 4,572 m)

Shipping and Storage Altitude

-2,000 to 40,000 feet (-610 to 12,192 m)

Operating Relative Humidity

15% to 95%, noncondensing

Shipping, Storage Relative Humidity

15% to 95%, noncondensing

Shock

50 g

Vibration

Random Vibration, 0.02g

2

/Hz from 10 to 300 Hz, ramping down to 0.002g

2

/Hz

at 500 Hz. Operating 1 hour per axis,
3 hours per test.

Water Resistance

For ordinary, indoor locations only.

Characteristic

Specification

Universal power adapter, 503-0054-00, 503-0093-XX

Length

5.0 in (12.7 cm)

Width

3.6 in (9.1 cm)

Height

3.1 in (7.9 cm)

Weight

3.1 lb (1.4 kg)

Rated Input

100V-120V ac, 500 mA, 50/60 Hz

Rated Fuses

T800 mA/250V, Time-Delay, 5x20mm

Rated Output (Continuous)

16-24V dc, 25 VA

Additional Features

Detachable power cord, pilot light

Universal power adapter, 503-0054-01, 503-0092-XX

Length

5.0 in (12.7 cm)

Width

3.6 in (9.1 cm)

Height

3.1 in (7.9 cm)

Weight

3.1 lb (1.4 kg)

Rated Input

200V-240V ac, 250 mA, 50/60 Hz

Rated Fuses

T400 mA/250V, Time-Delay, 5 x 20mm

Rated Output (Continuous)

16-24V dc, 25 VA

Additional Features

Detachable power cord, pilot light

Power Adapter

connector

+

-

Advertising