Welch Allyn Vital Signs Monitor, Propaq Encore - Quick Reference Guide User Manual

Page 8

Advertising
background image

2

General information

Welch Allyn Propaq Encore Vital Signs Monitor

WARNING During defibrillation, keep the discharge paddles away from ECG and
other electrodes, as well as other conductive parts in contact with the patient.
Avoid contact with any accessories connected to the Propaq’s left side panel.

WARNING To ensure patient safety, the conductive parts of the ECG electrodes
(including associated connectors) and other patient-applied parts should not
contact other conductive parts, including earth ground, at any time.

WARNING Do not operate this product in the presence of flammable
anesthetics or other flammable substance in combination with air, oxygen-
enriched environments, or nitrous oxide; explosion can result.

WARNING Within certain governmental jurisdictions, all interconnected
accessory equipment must be labeled by an approved testing laboratory. After
interconnection with accessory equipment, risk (leakage) current and grounding
requirements must be maintained.

WARNING To ensure conformance to risk (leakage) current requirements when
operating from an ac mains power source, use only a Welch Allyn

®

503-0054

series power adapter.

WARNING This monitor should only be repaired by qualified service personnel.
The operator should not attempt to open the monitor case or perform any
maintenance on the monitor except for procedures explicitly described in this
manual that can be performed by operators such as inspection and cleaning.

WARNING To ensure patient safety, use only accessories approved by Welch
Allyn. Visit

www.welchallyn.com

. The use of any other accessories can result in

inaccurate patient data, can damage the equipment, and can void your product
warranty.

WARNING Always use accessories according to the standards of your facility
and according to the manufacturer's directions for use.

WARNING Safe interconnection between the Propaq monitor and other devices
must comply with applicable medical systems safety standards such as IEC
60601-1-1. Within certain governmental jurisdictions, all interconnected accessory
equipment must be labeled by an approved testing laboratory. After
interconnection with accessory equipment, risk (leakage) current and grounding
requirements must be maintained.

WARNING As with all medical equipment, carefully route the patient cabling to
reduce the possibility of patient entanglement or strangulation.

WARNING A product that has been dropped or severely abused should be
checked by qualified service personnel to verify proper operation and acceptable
risk (leakage) current values.

WARNING The pulse oximetry channel should NOT be used as an apnea
monitor.

WARNING Do not use the pulse oximeter as a replacement or substitute for
ECG-based arrhythmia analysis.

Advertising