Replacing the fuse, Monitor functions based on battery voltage, Checking battery voltage – Welch Allyn Vital Signs Monitor, Propaq Encore - Quick Reference Guide User Manual

Page 98

Advertising
background image

92

Power sources

Welch Allyn Propaq Encore Vital Signs Monitor

For monitors without the Expansion Module or SpO

2

option, typical monitor operating

time is about 2 hours at 25° C for a new, fully-charged battery. This is when all patient
channels are active and measurements are taken every 15 minutes.

For monitors without the Expansion Module but with the SpO

2

option, operating time is

about 5 hours.

For monitors with the Expansion Module and printer, SpO

2

and CO

2

options, typical

operating time is about 3 hours under the above conditions when print strips are
generated every 15 minutes.

Monitor functions based on battery voltage

As battery voltage drops during extended monitor battery operation, error messages are
displayed and monitor functions are discontinued in order of priority.

Checking battery voltage

The Propaq’s battery voltage is displayed on the initial powerup screen. The battery
voltage is also displayed with the Settings window.

Replacing the fuse

The Propaq Encore is protected against power surges by a 3 Ampere fuse, which can
easily be replaced in the right side panel. Fuse replacement should only be performed by
a qualified service person.

1.

Disconnect the Propaq Encore from the patient.

2.

Disconnect the ac power adapter from the Propaq Encore’s power input connector.

3.

Using a small screwdriver or similar device, unscrew the fuse carrier by turning it
counterclockwise.

4.

Remove the fuse holder and replace the fuse with a 3 Ampere, 2AG, 250V (fast or
slow acting) fuse. This fuse can be ordered from Welch Allyn or its service centers
using part number 503-0058-00.

Note

If the green Battery Charging lamp does not light when an ac adapter is
connected, this fuse may be blown.

3A

2AG

12-28V, 3A

Replaceable

Fuse