How temperature is displayed

Temperature is displayed as a numeric only, in a window at the top of the Propaq Encore
screen, in °C or °F. This area displays all temperature measurements (T1, T2,

T), one at a

time.

Temperature messages

The following messages can appear in an equipment alert window. A temperature caution
message will also appear in the temperature numeric window when one of these
messages appears (except PROBE NOT DETECTED).

PROBE NOT DETECTED. This message occurs when the Propaq Encore has
successfully measured temperature and a probe is then disconnected. Reconnect the
probe or acknowledge the equipment alert by pressing any menu button.

PROBE SHORT. Verify that the probe is properly inserted in the left side panel. If so,
replace probe.

CALIBRATION ERROR, TEMP DISABLED. This message appears when the Propaq
Encore has detected that it cannot accurately measure the temperature. The monitor
should be serviced.

Malfunction of the temperature probes may result in inaccurate readings. Confirm
suspect readings.