Rr/br apnea – Welch Allyn Vital Signs Monitor, Propaq Encore - Quick Reference Guide User Manual

Page 69

Advertising
background image

Reference Guide

Patient monitoring

63

6.

Press RESPONSE to select either NORMAL, SLOW, or FAST.

The FAST setting is recommended where a sudden step change in ETCO

2

is of

concern, such as that induced by an air embolus in certain neurosurgical procedures.

A SLOW response will decrease ETCO

2

false alarms when breath morphology varies

considerably from one breath to the next.

The default setting is NORMAL.

z

7.

Set alarm limits for ETCO

2

, INCO

2

, and Breath Rate.

Motion artifact or other factors can cause false RR/BR alarms. To help minimize false
alarms, the Propaq delays or “holds off” triggering a RR/BR alarm for 5 seconds. After
this holdoff period begins, if the Propaq detects that the patient’s RR/BR vital sign has
returned to acceptable limits, the Propaq cancels the alarm holdoff. The next time a
RR/BR limit is violated, the Propaq starts a new 5-second RR/BR alarm holdoff period.

8.

Set the alarm limit for Apnea Delay—this is the maximum time allowed between two
successive breaths before an Apnea alarm occurs.

After the first breath has been detected, the Apnea alarm limit setting will be
automatically turned on for as long as the CO

2

channel is active. The RR/BR and

Apnea Alarm Limit window is shown below. STAT SET does not affect the Apnea
limit.

Note

If the ETCO

2

value is displayed as + + +, verify calibration against a known

reference gas. If the sensor calibration is not accurate, return it to Welch Allyn for
service.

Response Time
Setting

Sampling Period

Typical Indications for Use

FAST

15 seconds

During neuroanesthesia

NORMAL

30 seconds

During routine use

SLOW

45 seconds

To decrease ET

CO

2

false alarms

WARNING For patient safety, it is recommended that the Breath Rate alarm
limits always be turned on and set appropriately.

mmHg

NEXT

NEXT PAGE

UP

MCO

2

BR

SpO

2

PA

ALARM LIMITS

35/ 18

( 85)

45 12

92

DOWN

Br/M: 30 5

DELAY:

15

secs

UPPER LOWER

RR/BR

APNEA

ON/OFF