Reference Guide

Patient monitoring

47

NIBP Displayed in Waveform Window

Set the NIBP alarm limits

Set the alarm limits according to your hospital's standards.

NIBP display default settings

To select which vital sign waveforms are displayed, press

MAIN MENU, SETUP, WAVE SEL

to

display the Wave Select window.

To enable the display of a waveform or large NIBP numerics, select ON (ECG cannot be
set to OFF). The monitor displays the first three active waveforms set to ON in the order
of priority listed in the Wave Select window. If NIBP is ON and only one or two other
waveforms are ON and active, the monitor displays large NIBP numerics in a waveform
window.

mmHg

NIBP

SETUP

INV PRS

SpO

2

/CO

2

ECG/RESP

MCO

2

BR

HR

T

1

NIBP

18:45:28

III

1

mV

cm

MON

N

I

B

P

60
30
0

102.3

F

mmHg TIME SINCE READING: 0 min

35

135/ 81

(103)

12

80

NIBP

C

O

2

M

(103)

135

81

SpO

2

99

mmHg

NIBP

SETUP

INV PRS

SpO

2

/CO

2

ECG/RESP

MCO

2

BR

HR

T

1

NIBP

19:02:28

III

1

mV

cm

MON

N

I

B

P

60
30
0

102.3

F

SYS DIA MEAN

135 81 (103)

mmHg TIME SINCE READING: 17 min

35 12

80

NIBP

18:45

C

O

2

M

135/

81

(103)

SpO

2

99

NIBP TREND

TIME

BR

SYS/DIA-MEAN

HR/PR

BPM

Br/M

HH:MM
10:35

9:46

9:31

9:20

OFF

OFF

OFF

OFF

106

88

93

91

OFF

OFF

OFF

OFF

135/ 81 103

142/ 83 110

145/ 87 112

126/ 85 91

SpO2
%

mmHg

PR

T

1

NIBP

10:45:28

N

I

B

P

102.3

F

mmHg TIME SINCE READING: 10 min

106

NIBP

PRINT

OXYCRG

NXT TRND

135/ 81

(103)

(103)

135

81

The numerics are shown in large
characters for 16 minutes for each new
measurement taken...

By turning on NIBP in the wave select
window, the NIBP numerics can be
displayed in a waveform window.

... and then they change to the smaller
characters for 44 minutes. The numerics
are removed after 60 minutes.

If NIBP is the only vital sign being
monitored, the numerics are displayed in a
waveform window above a trend window.

WAVE SELECT

mmHg

NEXT

PREV MENU

INSERV

ON/OFF

MCO

2

BR

SpO

2

PA

SETUP

ECG : ON RESP : OFF
ART :

ON

SpO2 : ON

PA : OFF NIBP : ON
CO

2

: OFF

35/ 18

( 85)

35 12

92

Advertising