Factory default settings, Factory default, Settings – Welch Allyn Vital Signs Monitor, Propaq Encore - Quick Reference Guide User Manual

Page 124

Advertising
background image

118

Specifications

Welch Allyn Propaq Encore Vital Signs Monitor

Factory default settings

Characteristic

Specification

Date

a

MO/DA/YR, DA.MO.YR, or YR/MO/DA

Decimal

a

. (Period)

HR/PR Sweep

25 mm/sec

RR/BR Sweep

6.25 mm/sec

Alarm Tone

MEDIUM

HR/PR TONE

LOW

HR/PR SOURCE

ECG

RR/BR Source

CO

2

if ON or ECG (if CO

2

OFF)

Patient Mode

Adult

ECG Bandwidth

Monitor

ECG Size

1 mV/cm

ECG Lead

II

ECG Filter

a

60 Hz

ECG Pacer

ON

RESP size

2X

RESP lead

Ld2

RESP sweep

6.25 mm/sec

RESP ON/OFF

ON

RESP window

ON

IBP Range

0 to 180 mmHg

IBP Rescale

0 to 140 mmHg

IBP Mode

RESCALE

Invasive Pressure Formats

Label dependent

NIBP Mode

MANUAL

NIBP Auto Time

15 min

SpO

2

SIZE

2x

SpO

2

C-LOCK

OFF

SpO

2

Response

NORMAL

TEMP F/C

a

Celsius

CO

2

Range

0 to 60 mmHg

CO

2

Sweep

6.25 mm/sec

CO

2

Response

NORMAL

CO

2

Units

a

mmHg

CO

2

Gas Compensation

OFF

Sidestream CO

2

Flow Rate

Adult: 90 ml/minute
Ped: 90 ml/minute
Neonate: 90 ml/minute
(The flow rate cannot be programmed to a different value in a Custom Patient Mode.
See Custom Patient Modes.)

Display Wave Select

All waves are on except NIBP

Trend Group

NIBP

Alarm Limits

All are ON except P2

HR Limits

Adult: 50, 120 beats per minute
Ped: 50, 150 beats per minute
Neonate: 100, 200 beats per minute

NIBP Limits - Systolic

Adult: 75, 220 mmHg
Ped: 75, 145 mmHg
Neonate: 50, 100 mmHg

NIBP Limits - Diastolic

Adult: 35, 110 mmHg
Ped: 35, 100 mmHg
Neonate: 30, 70 mmHg

NIBP Limits - Mean

Adult: 50, 120 mmHg
Ped: 50, 110 mmHg
Neonate: 35, 80 mmHg

P1, P2 Limits - Systolic

Adult: 75, 220 mmHg
Ped: 75, 145 mmHg
Neonate: 50, 100 mmHg