HR/PR TONE

Sets the Heart Tone volume to HIGH, MEDIUM, LOW, or OFF.

PATIENT MODE Pressing CHANGE in this selection displays the following Patient

Mode window:

If you press NEONATAL, PED, or ADULT, a confirmation window
appears, requiring you to confirm your selection:

Note

Whenever you change the patient mode, the alarm limit settings are
automatically changed to the defaults for that mode. If Custom
settings have been set for that mode, the defaults are the Custom
mode settings. If no Custom settings have been set, the defaults are
the Factory Mode settings. See page 23 for more information about
patient modes.

If you change the patient mode, the CO

2

alarm limits in the new

mode might vary slightly from the originally-programmed CO

2

limits

for the new mode. Check the CO

2

alarm limits.

If you press SETUP in the previous Patient Mode window, the Mode
Setup window appears. This allows you to set custom patient
modes and powerup defaults as described on page 23.

SETUP

BATTERY: 9.2 VOLTS

mmHg

NEONATAL

PREV MENU

ADULT

PED

MCO

2

BR

SpO

2

PA

PATIENT MODE
SELECT PATIENT MODE BASED ON AGE:

NEO : < 44 WEEKS GEST. AGE
PED : > 44 WEEKS GEST. AGE,
< 9 YEARS
ADULT: > 9 YEARS

35/ 18

( 85)

35 12

92

mmHg

YES

NO

MCO

2

BR

SpO

2

PA

PATIENT MODE

CHANGING SETTINGS AND MODE

35/ 18

( 85)

35 12

92

ARE YOU SURE?

TO:

ADULT ALARM LIMITS WILL BE SET

CUSTOM: ADULT