Nibp trends, Displaying trends, Nibp trends displaying trends – Welch Allyn Vital Signs Monitor, Propaq Encore - Quick Reference Guide User Manual

Page 84: Selecting a trend (nxt trnd)

Advertising
background image

78

Trends

Welch Allyn Propaq Encore Vital Signs Monitor

The following is programmable: which trend window comes up first. The NIBP trend
window is the factory default.

To clear trend data before connecting a new patient, the monitor power must be cycled.
This will prevent the trend data of a previous patient from being attributed to the new
patient.

NIBP trends

A maximum of 128 NIBP readings are collected (up to 8 hours). NIBP is not measured
continuously like other vital sign parameters. The numerics on this trend are captured at
the time of the NIBP reading.

If an error number (ERR# x) is listed in an NIBP TREND printout or display, it indicates that
an NIBP equipment alert occurred. See

“NIBP messages”

on page 48 for NIBP alert error

numbers and definitions.

Displaying trends

Displayed trends show the last five hours of data. Trends are displayed if you are at the
main menu and have only one waveform turned on in the wave selection window. Trends
are also displayed when you press the TRENDS button in the first Setup Menu.

Selecting a trend (NXT TRND)

Select the trend you want displayed by pressing the NXT TRND button. The trend is
identified by a label at the top of the table.

mmHg

MCO

2

BR

SpO

2

35/ 18

( 85)

35 12

92

PRINT

OXYCRG

NXT TRND

NIBP TREND

TIME

BR

SYS/DIA-MEAN

HR/PR

BPM

Br/M

HH:MM
10:01

9:46

9:31

9:20

98

SRCH

OFF

100

101

88

93

91

11

10

5

11

139/ 74 106

142/ 83 110

145/ 87 112

126/ 85 91

SpO2
%

mmHg

PA

This symbol indicates that the NIBP reading was taken in
the presence of high motion artifact while monitoring
ECG. Artifact can affect accuracy. To help reduce artifact,
see

“Place and connect the cuff”

on page 44.

When the SEARCH message appears in an NIBP TREND
display or printout, it indicates that the monitor was not
able to complete an NIBP measurement during that
period.

P1 TREND

P2 TREND

NIBP TREND

RESP TREND

TEMP TREND

Time

Time

Time

Time

Time

HR/PR

HR/PR

HR/PR

HR/PR

HR/PR

SpO

2

SpO

2

SpO

2

SpO

2

SpO

2

SYS/DIA-Mean

SYS/DIA-Mean

SYS/DIA-Mean

RR or BR

T1

RR or BR

RR or BR

RR or BR

ETCO

2

T2

INCO

2

T