Printer functions

Press SETUP, MORE, PRINTER to display the printer menu and setup window.

NEXT

Selects the next setting in the status window.

CHANGE

Changes the currently selected display setting.

PR TREND

Prints all trends turned on in the Printer Trend Select Window.

MORE

Pressing the MORE button displays another menu and status
window.

PREV MENU

Returns you to the previous menu.

CONTINUOUS

Sets the print speed for real time (continuous) measurements to
6.25, 12.5, or 25 mm/sec. This sets the print speed for a printout
obtained by pressing the START/STOP button on the printer.

AUTO PRINT

Automatically prints 8 seconds of patient information every 15
minutes, 30 minutes, 1 hour, 2 hours, or 4 hours. This is the
latest patient information (real time). The print speed is
automatically set to 25 mm/sec.

ALARM PRINT

Automatically prints upon an alarm. The Propaq Encore prints 20
seconds of patient information. The first 12 seconds contain
information prior to the alarm. The print speed is automatically
set to 25 mm/sec.

NIBP TICKET

Automatically prints an NIBP Ticket when the measurement is
taken.

APNEA TICKET

When turned on, an Apnea Ticket is printed at the conclusion of
an apnea alarm and at the one-minute clock interval if the apnea
alarm does not cease.

OXYCRG ON
ALARM

When turned on, an oxycardiorespirogram will print if an HR/PR,
SpO

2

, or RR/BR alarm occurs. For more information on OxyCRG,

see

“OxyCRG”

on page 82.

mmHg

NEXT

MORE

PREV MENU

PR TREND

CHANGE

MCO

2

BR

SpO

2

PA

ART

HR

T

1

NIBP

18:45:28

III

1

mV

cm

MON

PRINTER

SETUP PAGE

CONTINUOUS : 25.0 mm/s
AUTO PRINT : 15 minutes
ALARM PRINT :

OFF

NIBP TICKET

: OFF

APNEA TICKET : OFF
OXYCRG ON ALARM: OFF

102.3

F

S

122

35/ 18

( 85)

35 12

92

122

58

80