Ä chapter 5.9.2 ‘opening the, Using the device – FRITSCH PULVERISETTE 7 premium line User Manual

Page 41

Advertising
background image

5.9.2 Opening the grinding bowl after a grinding operation

1.

Allow the bowl (1) to cool down.

2.

Release the pressure by turning the press-on sleeve (7) to open it.

NOTICE!
After releasing the pressure via the press-on sleeve,

this and the valve plugs (6) must be cleaned thor-

oughly before further grinding processes. If this is not

done, the tightness may be reduced due to residual

sample material.
After restarting and reaching the set speed, we rec-

ommend securing the press-on sleeve (7) and handle

(5) again.

3.

Turn the handle (5) to open it.

4.

If the lid insert (9) is sticking on the flat seal (3), apply a screw-

driver to the actuating pin (8) to remove the lid insert (9) from the

flat seal (3).

5.

Open the lid (2).

Using the device

- 41 -