6.2 Gassing lid with stainless steel, swagelok hose couplings

NOTICE!
Our workers perform a water bath test on all gassing lids.

The part to be tested is sealed, a pressure of 5.5 bar is

applied and it is immersed in a water bath. If there is a leak,

bubbles will develop. The air bubbles that develop within a

specific interval are evaluated by the worker/tester.
Only gassing lids with a leak rate of <10

-4

[mbar l/s] are

approved.

6.2.1 Scope of delivery / lid design

1 Compl. gassing lid handle

2 Compl. gassing lid fitting

3 Gassing valve connecting plug

4 Gassing valve quick-release coupling

5 Sealing ring 8x12x1mm, vulcanised fibre

6 Flat seal

7 Actuating pin

8 Gassing lid insert

9 Flat seal, 57.5x48x2, silicone

10 Hexagon socket wrench size 13