6.3.3 Configuration of transmitter ID and data transmission frequency

The measuring/transmission unit (12) is equipped with four configura-

tion switches for setting the transmitter ID and data transmission fre-

quency.

The switches illustrated are all OFF (actuator cams set to OFF).

6.3.3.1

Adjustment options of the transmitter detection

Upon delivery, all switches are set OFF. This means as identification that

it is cup 1 and the transmission frequency is 1 second.

ID

Switch 1

Switch 2

Transmitter no. 1

OFF

OFF

Transmitter no. 2

ON

OFF

Transmitter no. 3

OFF

ON

Transmitter no. 4

ON

ON

NOTICE!
For the operation of two or more EASY GTM systems, each

transmitter must have a separate ID.