KEYENCE DL-EP1 User Manual

Page 61

Advertising
background image

3-4 Message Communication

3-33

3

Commu

nic
ati

ng

wi

th t
he SK-1000

S

eries

- EtherNet/IP Compatible Network Unit DL-EP1 User’s Manual (SK-1000) -

Lock setting

Responses

The responses made when each service is used are as follows:

(1) Parameter read:

: Service code 0EH

(2) Parameter write:

: Service code 10H

(3) Operation command:

: Service code 4BH

(4) Batch lock setting:

: Service code 4CH

(5) Read of number of decimal places: : Service code 4DH

Setting

Description

0

Does not execute lock.

1

Executes key lock (inhibition of operations other than basic screen display

switching and [RESET] button).

General

Status

Command Type

Description

(1) (2) (3) (4) (5)

00H

Normal end

05H

The instance ID is out of range.

08H

-

-

-

The designated instance ID does not support this service or

cannot be executed.

09H

-

-

-

-

The written data is out of range.

0CH

-

-

-

•

Parameter write was executed for the function controlled

by cyclic communication.

•

An attempt to execute the operation command has failed.

Check if the sensor amplifier can execute the operation

command.

0EH

-

-

-

-

An attempt was made to write data to the attribute ID that

cannot be written or to the attribute ID that is in the unwritable

status.

10H

-

-

-

-

An attempt was made to read data from the attribute ID that

cannot be read or from the attribute ID that is in the

unreadable status.

13H

-

-

The service data size is less than the defined size.

The defined data size is stored in the additional status.

14H

-

The attribute ID is out of range.

16H

-

The sensor amplifier corresponding to the designated

instance ID is not connected.

1FH

•

This service is supported by the designated instance ID

but cannot be used by the designated attribute ID.

(Additional status: C350H)

•

In the current mode, the designated setting cannot be

written into the parameter.

(Additional status: C351H)

•

The DL-EP1 is now initializing communication. (Additional

status: C352H)

FEH

System error. Contact your nearest KEYENCE office.

Advertising