Contact, How to replace the dust seal, Amplifier part names and descriptions – KEYENCE GT2-500 Series User Manual

Page 6: How to replace the contact, How to mount the lift lever

Advertising
background image

6

GT2-500-M-E

Detaching the air tube

To detach the air tube, (1) press down on the release ring, (2) pull
out the tube.

● For GT2-A12K/A12/A32/A50 when an IP67 enclosure

rating is required

To maintain an IP67 enclosure rating for GT2-A12K/A12/A32/A50,
the following must be satisfied:

Firmly connect tubing with

2.5 mm inner diameter to the exhaust

port until it bottoms out.
Ensure that the tube end is free of dirt or water.

Tighten the screw of the exhaust valve with a torque of 0.5 to 0.6 N m.
If the screw is loose, the IP67 rating cannot be guaranteed.

● Adjustment of Spindle Movement Speed

To adjust the spindle movement speed, install a speed controller
between the cylinder and the air supply. Using OP-87970 is
recommended.

Contact

This section describes how to replace the contact, and how to mount
the lift lever.

How to replace the contact

While securing the spindle with the accessory key wrench, detach
the contact with pliers. Then attach a new contact.

How to replace the dust seal

Refer to the instruction manual included with the replacement dust seal
(OP-87932).

How to mount the lift lever

Mount the lift lever (OP-84397) between the spindle and the contact.
Secure the spindle with the lift lever and attach/detach the contact.

Amplifier part names and descriptions

Up to 5 sensor heads can be connected to 1 main unit/expansion
unit amplifier. Up to 2 expansion units can be added to 1 main unit.
Therefore, up to 15 sensor heads can be connected in total.

NOTICE

• Before detaching the tube, be sure to stop any air

flow.

• Press down on the release ring evenly from both

sides, and pull the tube out. Uneven pressure may
result in damage to the tube or damage to the
operation of the air cylinder.

Reference

By loosening the screw of the exhaust valve, you can
change the exhaust port angle.

NOTICE

• To further decrease spindle movement speed, use

a coil-shaped tube (OP-87986) or similar to
increase the distance between the air supply hole
and the speed controller.

• The speed controller will not operate if installed in

the reverse direction.

Release ring

(2)

(1)

Screw

Air tube

Air tube

Exhaust valve

Speed controller

Air supply

Indication symbol

Adjusting the speed.

NOTICE

• Detach the sensor head from the device or fixture

(metal plating, etc.) before replacing the contact.

• When applying pliers to the contact, be sure not to

rotate the main part and cover the contact with a cloth.

• Never apply tightening torque over 0.2 N m when

attaching a contact.

• Fix the roller contact (OP-77680) or the offset contact

(OP-77683) with the fixing nut in the same direction as in
actual use.

• Applying an adhesive, a thread locking agent, is

recommended to prevent the nut from getting loose.

• Position the roller contact carefully in the proper

direction. Care must be taken not to apply misdirected
force to the shaft.

Item

Explanation

Sensor head
connector

Connect to a sensor head cable with a sensor head attached.
Refer to "Connecting the amplifier" (page 3) for details.

DIP switch

Channels set to ON (enable) can recognize sensor heads.
The factory default setting for all switches is ON.
Switch channels that will not be used to OFF (disable).

Be sure to change DIP switch settings with the power off.
If the equipment is turned on while the DIP switch settings are
changed, setting changes will not be effective.
Changes are applied when the power is turned on.

Status light

Lit in green:Normal. Lights when the DIP switch is set to ON

(enabled), and the sensor heads are corrected properly.

Lit in red:

Error. Lights when the DIP switch is set to ON (enabled),
and there is an error.
See page 8 for information on checking errors and taking
corrective action.

Not lit:

Disabled Stays unlit when the DIP switch is set to OFF
(disabled).
If a sensor head is connected, it will not be automatically
recognised (the sensor head indicator will not light up).

Power cable

Supplies power. The cable is 2 m long.
Expansion units do not have power cables. Power is supplied to
expansion units through the main unit.

Spindle

Contact

Key wrench

Pliers

Cover the contact
with a cloth

GT2-500 (Main unit)

GT2-550 (Expansion unit)

Advertising