oPerAting the ultrAWAVe series in leVel APPlicAtions

introduction

The UltraWave Series is designed to get you up and running quickly once the system is installed. The controller
guides you through the Startup Procedure. See

Figure 11: Startup Procedure Diagram

. During the Startup

Procedure, you will be asked to input a number of parameters, including:

• Tank Height: The distance from the full point to the point where the vessel is considered empty. Material

level changes within this range are displayed by the UltraWave Series controller. See

Figures 12 and 13

.

Note: The maximum distance a transducer can measure depends on its frequency and the type of liquid in the
vessel. Entering a value for Operating Span greater than the transducer’s capability may cause erratic reading.

Note: See Calibration Menu for a description of the full point.

# of Transducers

> 8 pts

Up to 8 level

channels per

controller

Level

Chapter

4

Flow

Chapter

5

Sensor 1 App?

* Level

Figure 11: Startup Procedure Diagram

stArt-uP

When performing the Startup Procedure, use backspace if you make an error. The parameters can also be
corrected in the Calibration Menu after you complete the Startup, or the entire Startup Procedure can be
repeated by defaulting the channel in the Service Menu. Perform the Startup Procedure, as follows:

1. Apply power to the unit. If there are Scanner PCB(s) installed, the display is shown below. If not, go to

step 2.

2. Use the keypad to input the number of transducers you want to set up. Press ENTER to save the value in

memory.

3. Press ENTER to select Level.

Sensor 1 App?

* Level

40