oPerAting the ultrAWAVe series in leVel APPlicAtions
introduction
The UltraWave Series is designed to get you up and running quickly once the system is installed. The controller
guides you through the Startup Procedure. See
Figure 11: Startup Procedure Diagram
. During the Startup
Procedure, you will be asked to input a number of parameters, including:
• Tank Height: The distance from the full point to the point where the vessel is considered empty. Material
level changes within this range are displayed by the UltraWave Series controller. See
Figures 12 and 13
.
Note: The maximum distance a transducer can measure depends on its frequency and the type of liquid in the
vessel. Entering a value for Operating Span greater than the transducer’s capability may cause erratic reading.
Note: See Calibration Menu for a description of the full point.
# of Transducers
> 8 pts
Up to 8 level
channels per
controller
Level
Chapter
4
Flow
Chapter
5
Sensor 1 App?
* Level
Figure 11: Startup Procedure Diagram
stArt-uP
When performing the Startup Procedure, use backspace if you make an error. The parameters can also be
corrected in the Calibration Menu after you complete the Startup, or the entire Startup Procedure can be
repeated by defaulting the channel in the Service Menu. Perform the Startup Procedure, as follows:
1. Apply power to the unit. If there are Scanner PCB(s) installed, the display is shown below. If not, go to
step 2.
2. Use the keypad to input the number of transducers you want to set up. Press ENTER to save the value in
memory.
3. Press ENTER to select Level.