Figure 17: Flow Calibration Menu

Prim: Select primary element
type and size

Zero: Calibrate the system

Sim: Simulate head to check
corresponding flow

Unit: Set up standard units and
format for flow rate and total
flow

LHCO: Set low head cutoff

ReSet: Reset minimum flow,
maximum flow and total flow

Avg: Establish number of
reading to average for display

Temp: Sets up temperature
probe parameters

cAlibrAtion Menu

This section provides a general overview of the Calibration Menu. The Startup Procedure got your system up and
running. The Calibration Menu is used to enter setup and calibration parameters if your application requires a more
detailed set up, and/or to modify parameters you originally entered during startup.

21 PriM (PriMAry eleMent)

The primary element type and size were selected during the Startup Procedure. The Prim Menu allows you to
display and modify the primary element type and size. The primary element type choices are:

• Flum (flume) — Parshall, Palmer-Bowlus, Leopold-Lagco, Trapezoidal, and Other

• Weir — Rectangular (with and without end contractions), V-Notch, Cipolletti, and Other

• Nozz (nozzle) — Kennison

22 zero

The Zero Menu allows you to display and modify the Auto or Manual calibration parameters.

The Auto parameters are:

• Max Head — The maximum head for your primary element, available from the flow chart supplied by the

primary element manufacturer.

48

97-1163-01 Rev. D

www.kistlermorse.com