‘t’ coMMAnd — request totAlizer dAtA (floW APPlicAtions only)

This command is sent by the master to request the controller to send the totalizer data. For this example, the
master requests the totalizer data from a controller channel at address 01, which is set up with standard totalizer
format of XXXX.X and standard totalizer units of cubic feet. The controller returns a total flow of 2753 (275.3 cubic
feet) and a checksum of A5.

Request from Master

Response from Controller

>aatssr

Ahhhhhhhhhhssr

Format

>01tD

5r A0400000AC1A5r

Example

Note 1. 0=CF, 1=GL, 2=MG
Note 2. Format: 0=XXXXXX,1=XXXX00, 2=XXXXX0, 4=XXXX.X, 5=XXX.XX, 6=XX.XXX, 9=X.XXXX, B=XXX000
Note 3. 32 bit totalizer data in hexadecimal; 00000AC1 is the hexadecimal equivalent of 2753.

‘rd’ coMMAnd — request rAW tArget

This command is sent by the master to request the controller to send the raw target. For this example, the master
requests the raw target from a controller channel at address 01, which returns a raw target of 25939 (259.39
inches) and a checksum of 6C.

Request from Master

Response from controller

>aaRDssr

Adddddddssr

Format

>01RDF7r

A00259396Cr

Example

note 1

note 3

note 2

note 1

note 3

note 2

VII. MAINTENANCE

There is no preventative maintenance required for this unit. However, it is recommended that the face of the
sensor be cleaned every three months. The sensor should be wiped off with a clean towel.