introduction

Refer below to the block diagram of the initial part of the Startup Procedure, and

Figure 15

for the remainder of

the Startup Procedure. The UltraWave Series is designed to get you up and running quickly once the system is
installed. The UltraWave Series controller guides you through the initial Startup Procedure. During the Startup
Procedure, you will be asked to input a number of parameters, including the following (

See Figure 16

):

• Maximum Head — The maximum head for your primary element is available from the flow chart supplied

by the primary element manufacturer.

• Zero Distance — If doing a Manual Calibration during Startup, measure the zero flow distance, the

distance from the face of the transducer to the bottom of the channel.

• Current Head — If doing an Auto Calibration during Startup, measure the current head height just before

you input the value to perform an accurate calibration.

# of Transducers

> 8 pts

Up to 4

channels per

controller

Level

Chapter

4

Flow

Chapter

5

Sensor 1 App?

Flow

Note: The maximum distance a transducer can measure depends on the type of material being measured. Do
not enter a value for Zero Distance greater than the transducer’s capability. Zero Distance minus Max head must
be greater than 12”.