Kistler-Morse KM ultrawave User Manual

Page 45

Advertising
background image

4. Press ENTER and the display shows:

Transducer Freq

* 43 KHz

The model number of the transducer indicates the nominal frequency of operation: SC43PT = 43KHz,
SC22PT = 22KHz. Refer to the model number of the transducer for the proper frequency selection.

Use the up/down arrow keys to select the desired frequency and press ENTER.

5. Press the up/down arrows to select the desired units. Press ENTER to save.

Operating Units?

* Feet

6. Press the up/down arrows to select the decimal place format. Press ENTER to save.

Decimal Place?

* XXXX.X

Note: Make sure transducer is perpendicular to liquid level.

7. Press the up/down arrows to select standpipe. Press ENTER to save.

Standpipe?

* No

Note: Only if no temperature probe is detected.

8. Use the numeric keys to enter the temperature. Press ENTER to save.

Set Temperature

> 67.8

9. Select YES to continue through the menus as above for the second sensor, selecting vessel parameters

and identification. Select NO to discontinue the setup of the remaining channels. See the Service Menu
to copy all parameters to the remaining channels or cycle power to start again. Once complete, you will
be given a chance to auto assign inputs and outputs. See page 18.

Vessel Height

Call More Chans?

* Yes

Note:

Goes on to 10 when Call More Chans is answered NO.

10. Auto assigns 4-20mA output.

Auto 4-20mA

* No

Notes:

1. Full Point + Span must be less than the Maximum Range for the transducer, or the UltraWave Series will not
be able to detect material levels near the bottom of the vessel.

41