stArt-uP

When performing the Startup Procedure, do not worry if you make an error. The parameters can be corrected
after you complete the startup, in the Calibration Menu,

Figure 17

, or the entire Startup can be repeated by

defaulting the channel in the Service Menu. Perform the Startup Procedure, as follows:

1. Apply power to the unit. If there are Scanner PCB(s) installed, the display shows (If not, go to Step 2):

Use the keypad to input the number of transducers you want to set up. Press ENTER to save the value in
memory.

# of Transducers

Up/Down/Enter

2. Press the arrow keys to select Flow and press ENTER. If a temperature compensation sensor is not

connected, a warning displays stating that one was not detected. Press any key to continue.

Sensor 1 App

* Flow

3. Press the Enter Key to select Inches or scroll with the arrow keys to select Centimeters.

Operating Units?

* Inches

4. Press the Enter Key to select Flume, or the arrow keys to select Weir or Nozzle as the primary element.

Primary Element

Flume

Note:

If you select Flume, continue with Step 5. If you select Weir, go to Step 7. If

you select Nozzle, go to Step 8.

5. Flume — The five flume selections are: Parshall, Palmer-Bowlus, Leopold-Lagco, Trapezoid, and Other. If

you selected Flume in Step 5, the display shows:

Flume Select

Parshall

Press the arrow keys to access the remaining options of the Flume Menu. Press the Enter Key for your

desired option.

6. For each option, there are sub-menus. Scroll through the options and make your selection. Once

selected, press ENTER and go to step 9: Select Flow Type.

Parshall Flume

1”

7. Weir — The four Weir selections are: Rectangular, V-Notch, Cipolletti, and Other.

Weir Select

Rectangular

45

97-1163-01 Rev. D