LogTag Analyzer User Guide (1.9)

not:

What the LogTag recorded

What the LogTag indicated

2 readings out-of-range, return to normal,

2 readings out-of-range, return to normal,

2 readings out-of-range, return to normal

Neither alert condition is met.

The maximum number of consecutive

out-of-spec readings is only 2 (5 required.)

The total out-of-range events equals 2 + 2 + 2

= 6. The alert condition is 7 total, so the alert
does not go off.

No Flashing Red Alert Indicator

4 readings out-of-range, return to normal,

4 readings out-of-range

The first alert condition is not met, as the the

maximum number of consecutive alert condition

readings is only 4 (5 required.)

The second alert condition however IS met

since the total alert condition readings equals 4
+ 4 = 8. The alert condition is 7 total. so the

alert DOES go off.

Flashing Red Alert Indicator

4 readings out-of-range, return to normal,

2 readings out-of-range

Neither alert condition is met.

The maximum number of consecutive alert

condition readings is only 4 (5 required.)

The total number of alert condition readings
equals 4 + 2 = 6. The alert condition is 7 total,

so the alert does not go off.

No Flashing Red Alert Indicator

6 readings out-of-range, return to normal

The second alert condition is not met, as the

total out-of-range events equals 6 (7 required.)
The maximum number of consecutive alert

condition readings however is 6, the alert

condition is 5 consecutive readings, so the alert

DOES go off.

Flashing Red Alert Indicator

Please note following special conditions:

1. If you wish an alert to be indicated on the first reading that is recorded out-of-spec,
place a tick in the box next to "Enable the Alert (Red) indicator when" and remove the