Troubleshooting

Section 6

YSI Incorporated

Environmental Monitoring Systems Operations Manual

6-

4

Water in probe connector

Dry connector; reinstall probe

2.3.2

Probe has been damaged

Replace probe

9

Calibration solutions out of spec or
contaminated with other solution

Use new calibration solutions

C

ORP fails Zobell check

Take into account temperature
dependence of Zobell solution
readings

5.4

Internal failure

Return sonde for service

9


Depth unstable or
inaccurate

Depth sensor has not been zeroed

Follow depth zero procedure

2.6.1

Depth sensor access hole is
obstructed

Follow depth cleaning procedure

2.10.2

Depth sensor has been damaged

Return sonde for service

9

Internal failure

Return sonde for service

9

Conductivity unstable
or inaccurate. Error
messages appear during
calibration.

Conductivity improperly
calibrated.

Follow cal procedure

2.6.1

Conductivity probe requires
cleaning

Follow cleaning procedure

2.10.2

Conductivity probe damaged

Replace probe

9

Calibration solution out of spec or
contaminated

Use new calibration solution

C

Internal failure

Return sonde for service

8

Calibration solution or sample
does not cover entire sensor.

Immerse sensor fully.

2.6.1

Installed probe has no
reading

The sensor has been disabled

Enable sensor

2.9.7

Water in probe connector

Dry connector; reinstall probe

2.3.2

Probe has been damaged

Replace the 6560 probe

9

Report output improperly set up

Set up report output

2.9.6

Internal failure

Return sonde for service.

9

Temperature, unstable
or inaccurate

Water in connector

Dry connector; reinstall probe

2.3.2

Probe has been damaged

Replace the 6560 probe

9

Optical probe (turbidity,
chlorophyll, rhodamine
WT, BGA-PC, or BGA-
PE) unstable or
inaccurate. Error
messages appear during
calibration

Probe requires cleaning.

Follow probe cleaning procedure

2.10.2

Probe requires calibration

Follow cal procedures

2.6.1

Probe has been damaged

Replace probe

9

Water in probe connector

Dry connector; reinstall probe

2.3.2

Calibration solutions out of spec

Use new calibration solutions

C

Wiper is not turning or is not
synchronized.

Activate wiper. Assure rotation.
Make sure setscrew is tight.

2.10.2

Wiper is fouled or damaged.

Clean or replace wiper.

2.10.2

Internal failure.

Return probe for service.

9