YSI 600LS - User Manual User Manual

Page 373

Advertising
background image

Specifications

Appendix O

YSI Incorporated

Environmental Monitoring Systems Operations Manual

O-10

Non-Vented Level - Shallow
Sensor Type..............Stainless steel strain gauge
Range........................0 to 30 ft (9.1 m)
Accuracy ................ +/- 0.06 ft (0.018 m)
Resolution.................0.001 ft (0.001 m)
Temperature Range -5 to 45 C

Vented Level - Shallow
Sensor Type..............Stainless steel strain gauge
Range........................0 to 30 ft (9.1 m)
Accuracy, 0-30ft ......+/- 0.01 ft (0.003 m)
Resolution.................0.001 ft (0.001 m)
Temperature Range -5 to 45 C

Temperature
Sensor Type...........Thermistor
Range.....................-5 to 50

C

Accuracy................+/- 0.15

C

Resolution..............0.01

C

Depth……………200 meters

Rapid Pulse Dissolved Oxygen, % saturation
Sensor Type...........Rapid Pulse - Clark type, polarographic
Range.....................0 to 500 % air saturation
Accuracy................0-200 % air saturation, +/- 2 % of the reading or 2 % air saturation, whichever is greater
200-500 % air saturation, +/- 6 % of reading
Resolution..............0.1 % air saturation
Temperature Range -5 to 50 C
Depth……………200 meters

Rapid Pulse Dissolved Oxygen, mg/L
(Calculated from % air saturation, temperature and salinity)
Sensor Type.......... Rapid Pulse - Clark type, polarographic
Range.....................0 to 50 mg/L
Accuracy................0 to 20 mg/L, +/- 2 % of the reading or 0.2 mg/L, whichever is greater

20 to 50 mg/L, +/- 6 % of the reading

Resolution..............0.01 mg/L
Temperature Range -5 to 50 C
Depth……………200 meters

ROX Optical Dissolved Oxygen, % saturation
Sensor Type...........Optical, Luminescence Lifetime
Range.....................0 to 500 % air saturation
Accuracy ...............0-200 % air saturation, +/- 1 % of the reading or 1 % air saturation, whichever is greater
200-500 % air saturation, +/- 15 % of reading; Relative to Calibration Gases.
Resolution..............0.1 % air saturation
Temperature Range -5 to 50 C
Depth…………… 61 meters

ROX Optical Dissolved Oxygen, mg/L
(Calculated from % air saturation, temperature and salinity)
Sensor Type.......... Optical, Luminescence Lifetime
Range.....................0 to 50 mg/L
Accuracy................0 to 20 mg/L, +/- 1 % of the reading or 0.1 mg/L, whichever is greater

20 to 50 mg/L, +/- 15 % of the reading; Relative to Calibration Gases.

Resolution..............0.01 mg/L
Temperature Range -5 to 50 C
Depth…………… 61 meters