Sondes

Section 2

YSI Incorporated

Environmental Monitoring Systems Operations Manual

2-36

RECOMMENDED VOLUMES OF CALIBRATION REAGENTS


The approximate volumes of the reagents are specified below for both the upright and inverted orientations.
Note that the volume values are only estimates. The actual amount of calibrator solution required will
depend on how many and what type of other probes are installed in you sonde bulkhead.


Table 1A

6820V2-1 and 6920V2-1 Sondes with Standard Calibration Cup

*

Probe to Calibrate

Upright

Inverted

Conductivity

200ml

150ml

pH/ORP

125ml

175ml

ISE

125ml

175ml

All Optical Sensors

50ml

DO NOT CALIBRATE***


Table 1B

6820V2-1 and 6920V2-1 Sondes with Optional Extended Calibration Cup*

Probe to Calibrate

Upright

Inverted

Conductivity

320ml

150ml

pH/ORP

240ml

175ml

ISE

240ml

175ml

All Optical Sensors

225ml

DO NOT CALIBRATE***



Table 2

6820V2-2 and 6920V2-2 Sondes with Extended Calibration Cup

*

Probe to Calibrate

Upright

Inverted

Conductivity

310ml

150ml

pH/ORP

200ml

150ml

ISE

200ml

150ml

All Optical Sensors

225 ml

DO NOT CALIBRATE***



Table 3

600XL and 600XLM Sondes

Probe to Calibrate

Upright

Inverted

Conductivity

50ml

50ml

pH/ORP

25ml

50ml



Table 4

6600V2-2 Sonde with Short Calibration Cup and Long Cup in Parentheses*

Probe to Calibrate

Upright

Inverted

Conductivity

425ml (650ml)

250ml (250ml)

pH/ORP

300ml (500ml)

250ml (250ml))

ISE

300ml (500ml)

250ml (250ml)

All Optical Sensors

180ml (500ml)

DO NOT CALIBRATE***


Table 5

6600EDSV2-2 Sonde with Short Calibration Cup and Long Cup in Parentheses*

Probe to Calibrate

Upright

Inverted

Conductivity

275ml (520ml)

350ml (350ml)

pH/ORP

175ml (400ml)

350ml (350ml)

All Optical Sensors

225ml (420ml)

DO NOT CALIBRATE***