ADLINK PXI-2502 User Manual

Page 12

Advertising
background image

4

• Introduction

Analog Input (AI)

Number of channels: 4 single -ended for DAQ/PXI-2502, 8 single-ended
for DAQ/PXI-2501

AD converter: LTC1416

Max sampling rate: 400KS/s

Resolution: 14 bits

FIFO buffer size: 2K samples

Input range: Bipolar:

±

10V, unipolar: 0~10V

Over voltage protection: Continuous

±

35V maximum

Input impedance: 1G

| 6pF

Trigger mode: Pre -trigger, post-trigger, middle-trigger, and delay trigger

Data transfers: Programmed I/O, and bus-mastering DMA with scat-
ter/gather

Input coupling: DC

Offset error:

Before calibration:

±

40mV max

After calibration:

±

1mV max

Gain error:

Before calibration:

±

0.4 % of output max

After calibration:

±

1mV of output max

General Purpose Digital I/O (G.P. DIO)

Number of channels: 24 programmable Input/Output

Compatibility: TTL/CMOS

Input voltage:

Logic Low: VIL=0.8V max.; IIL=0.2mA max.

High: VIH=2.0V max.; IIH=0.02mA max

Output voltage:

Low: VOL=0.5 V max.; IOL=8mA max.

High: VOH=2.7V min; IOH=400

µ

A

Advertising
This manual is related to the following products: