Operation Theorem
• 45
4.5.2.4 High-Hysteresis analog trigger condition
Figure 4.5.5 shows the high-hysteresis analog trigger condition, the trigger 
signal asserts when the input analog signal level is higher than the 
High_Threshold voltage, where the hysteresis region is determined by the 
Low_Threshold voltage. 
Figure 4.5.5 High-Hysteresis analog trigger condition
4.5.2.5 Low-Hysteresis analog trigger condition
Figure 4.5.6 shows the low-hysteresis analog trigger cond ition, the trigger 
signal asserts when the input analog signal level is lower than the 
Low_Threshold voltage, where the hysteresis region is determined by the 
High_Threshold voltage. 
Figure 4.5.6 Low-Hysteresis analog trigger condition