Using a four-wire connection – ADLINK PCI-9524 User Manual
Page 50
 
34
Signal
Connections
ADLINK Technology Inc.
Copyright 2008
Figure 3-4: Connecting to a four-terminal load-cell transducer
using a four-wire connection
+Vex
+Vo
-Vex
-Vo
AIn+
AIn-
VEXn+
VEX_SENn+
VEXn-
VEX_SENn-
NC
NC
Load-cell Transducer
PCI-9524
It is recommended to enable the remote-sense function, and loop-
back the VEXn+/- to VEX_SENn+/- on the terminal board you're 
using when connecting to the transducer. A lengthy extension 
cable between PCI-9524 and the terminal board inevitably has 
some lead resistance that results in voltage drop; looping-back the 
excitation on the terminal board creates a six-wire connection and 
compensates for voltage drop.
A better approach is to use specially designed load-cell transduc-
ers having two additional SENSE terminals. The voltage difference 
across the bridge excitation junctions is fed back to the voltage 
excitation circuitry by two separate 'sense-wires', to further correct 
the voltage drops due to the resistance in the excitation wiring. Be 
sure to enable the remote-sense function to take full advantage of 
a six-wire connection. A typical six-wire connection is shown in 
Figure 3-5.