Using a six-wire connection – ADLINK PCI-9524 User Manual

Page 51

Advertising
background image

Signal Connections

35

ADLINK Technology Inc.

PCI-9524

Copyright 2008

User’s Manual

Figure 3-5: Connecting to a six-terminal load-cell transducer

using a six-wire connection

+Vex

+Vo

-Vex

-Vo

Load-cell Transducer

+Sense

-Sense

AIn+

AIn-

VEXn+

VEX_SENn+

VEXn-

VEX_SENn-

NC

NC

Whether to use a six-wire connection is dependent on the imped-
ance of the load-cell transducers you are using, length of the wir-
ing cable, wire-gauge inside the cable, and the required
measurement accuracy. We recommend you to use a six-wire
connection as the default connection method for high-accuracy
load-cell transducers.

To add remote-sense capability to a four-terminal load-cell trans-
ducer, simply run two separate sense-wires, and join them
together with the excitation wires at the transducer's excitation ter-
minals.