Figure 3-8: floating source and differential input – ADLINK PCI-9524 User Manual
Page 53
 
Signal Connections
37
ADLINK Technology Inc.
Copyright 2008
Figure 3-8 shows how to connect a floating signal source to PCI-
9524 in differential input mode. For floating signal sources, you 
need to add a resistor at each channel to provide a bias return 
path. The resistor value should be about 100 times the equivalent 
source impedance. If the source impedance is less than 100
Ω
, you
can simply connect the negative side of the signal to AIGND as 
well as the negative input of the Instrumentation Amplifier without 
any resistors. In differential input mode, less noise couples into the 
signal connections than in single-ended mode.
Figure 3-8: Floating source and differential input
AIn+
AIn-
n = 4, ..., 7
Ground 
Referenced 
Signal 
Source 
To A/D 
Converter 
AIGND
+
-
Instrumentation
Amplifier 
+
-
Input Multipexer