Rx frequency – EXFO FTB/IQS-88000 Series Power Blazer for FTB-2/2Pro/500/IQS-600 User Manual
Page 146

Test Setup - Test Configurator, Timer, and System
134
FTB/IQS-88000 Series
Frequency
Note: The following frequency statistics are available for each lane.
RX Frequency
Frequency (GHz) indicates the frequency of the input signal.
Offset (ppm) indicates the frequency offset between the standard rate
specification and the rate at the input signal.
Note: For both Frequency and Offset the following background colors are used.
Max. Negative Offset (ppm) indicates the maximum negative
frequency offset between the standard rate specification and the rate
from the received signal.
Max. Positive Offset (ppm) indicates the maximum positive
frequency offset between the standard rate specification and the rate
from the received signal.
Note: Refer to Interface on page 279 for more information on standard rate
specifications.
Background color
Description
Green
The frequency is in range.
Red
The frequency is out-of-range or LOC Lane. LOC is also
displayed.
Gray
Pending state.