Hanna Instruments HI 83214 User Manual

Page 20

Advertising
background image

20

SPECIFICATIONS

Range

0.00 to 3.00 mg/L (as NH

3

-N)

Resolution

0.01 mg/L

Accuracy

±0.10 mg/L or ±5 % of reading @ 25 °C, whichever is greater

Light Source

Tungsten lamp with narrow band interference filter @ 420 nm

Method

Adaptation of the

ASTM Manual of Water and Environmental Technology, D1426-92,

Nessler method. The reaction between ammonia and reagents causes a yellow tint in
the sample.

REQUIRED REAGENTS

Code

Description

Q.ty/test

Q.ty/set

HI 93764A-0*

Reagent Vial

1 vial

25 vials

HI 93764-0

Nessler Reagent

4 drops

1 bottle

*

Reagent Vial identification: A LR, white label.

Note: Store the unused vials in a cool and dark place.

REAGENT SET

HI 93764A-25

Reagents for 25 tests

For other accessories see page 68.

MEASUREMENT PROCEDURE

• Select the

Ammonia LR method following one of the procedures described in the “Method Selection”

section (see page 15).

Ammonia LR

AMMONIA LOW RANGE

• Remove the cap from a Reagent Vial.

• Add exactly 5.0 mL of sample to the vial, while keeping the

vial at a 45-degree angle.

• Replace the cap and mix by inverting the vial a couple of

times. This is the blank.

• Place the vial into the holder and push it completely down.