Hanna Instruments HI 83214 User Manual

Page 27

Advertising
background image

27

• Remove the cap and add the content of one packet of HI 93711-0 Total

Chlorine Reagent.

• Replace the cap and shake gently to mix for about 20 seconds. This is the

sample.

• Place the vial into the holder and push it completely down.

• Press

Timer and the display will show the countdown prior to the measurement and the message

Reaction Time”. Alternatively, wait for 2 minutes and 30 seconds and press Read. The instrument

will perform the reading.

• The instrument directly displays concentration in

mg/L of total chlorine.

INTERFERENCES

Interferences may be caused by:
Bromine (Br

2

)

Iodine (I

2

)

Oxidized forms of Chromium and Manganese
Ozone (O

3

)

Alkalinity above 250 mg/L CaCO

3

or acidity above 150 mg/L CaCO

3

will not reliably develop the full amount

of color or it may rapidly fade. To resolve this, neutralize the sample with diluted HCl or NaOH.
In case of water with hardness greater than 500 mg/L CaCO

3

, shake the sample for approximately

2 minutes after adding the powder reagent.

Chlorine, Total